36
Using ProLink
®
Software with Micro Motion
®
Transmitters
File Menu: Print
continued
Print setup
2. At the Add Transmitter Tag dialog box, use the Tag Name option
buttons to select a device. If a device other than the connected
transmitter [CurCon] is selected, enter the polling address or HART
tag name in the text box. A device cannot be added to the ticket
definition file if a device with the same tag name already exists in the
file.
3. Use the Transmitter Type option buttons to specify one of three
available transmitter types:
• Select RFT9739 to poll for parameters available from an
RFT9739. When using Modbus protocol, select this option for all
devices on the ticket.
• Select RFT9712/9729/9739 to poll under HART protocol for
parameters common to all three transmitter models.
• Select IFT9701 to poll for parameters available from an IFT9701
or IFT9703.
• Select Other HART v.5 or higher to poll for parameters available
from transmitters or other devices that have HART 5.0 or higher
revision software.
4. Click OK to add the device to the ticket.
To remove a device from the ticket:
in the Transmitter Connections
frame, select the device from the Tag Names list, then choose Remove.
When a device is removed from a ticket, the fields associated with the
device are also removed.
Fields on ticket
To add a parameter to a device on the ticket:
1. Choose Add in the Fields on Ticket frame. The Edit Field Tag
Parameter dialog box appears as shown in
Figure 4-4
.
2. Select a device from the Tag Name text box. Each device includes
the polling address or HART tag name, followed by the transmitter
type in brackets.
3. All the parameters available from the chosen device appear in the
Parameter list box. To add a parameter, select it, then click OK.
Figure 4-4. Edit Field Tag
Parameter dialog box