Micro Motion RFT9709 Manual de instrucciones Pagina 127

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 214
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 126
Using ProLink
®
Software with Micro Motion
®
Transmitters
113
Configure Menu: Transmitter Outputs
continued
Milliamp outputs
Configure Menu:
Transmitter Outputs
Configure Menu:
Transmitter Information
Configure Menu: EventsView Menu: Status
Configure Menu:
Characterize
Configure Menu:
Transmitter Variables
The IFT9701 4-20 mA output can indicate mass flow or volume flow.
To select variables indicated by the IFT9701 or RFT9703 mA output,
choose a mass or volume flow unit from the Configure Transmitter
Variables dialog box.
The flow unit that is chosen determines whether the IFT9701 4-20
mA output will represent mass flow or volume flow.
To choose a mass or volume flow unit, see
Section 9.2
(page 93).
The RFT9712 and RFT9729 4-20 mA output can indicate mass flow,
volume flow, density, or temperature.
To select variables indicated by the RFT9712 or RFT9729 mA output,
use the drop-down list box under Milliamp 1.
If Mass Flow is chosen, the type of flow (mass or volume) indicated
by the output is dependent on the flow unit of measure that was
selected in the Configure Transmitter Variables dialog box.
To choose a mass or volume flow unit of measure, see
Section 9.2
(page 93).
Range limits
Use the 4 mA and 20 mA text boxes to program the range of process
variables assigned to milliamp outputs. The range represents values of
the variable at 4 mA and 20 mA if the output produces a 4-20 mA
current, or at 0 mA and 20 mA if the output produces a 0-20 mA current.
Low Limit, High Limit, and Minimum Span, displayed at the bottom of the
dialog box, indicate sensor limits in the measurement units established
for milliamp output variables.
The value entered into the 4 mA text box must be lower than the
value entered into the 20 mA text box.
The value entered into the 20 mA text box must be higher than the
displayed low limit and lower than the displayed high limit.
The difference between the values entered into the text boxes must
be greater than the displayed minimum span.
Milliamp output flow
cutoffs for RFT9739 and
RFT9712/RFT9729
If the configuration applies to an RFT9739, RFT9712, or RFT9729, enter
flow cutoffs for milliamp outputs into the Cutoff text boxes. Flow cutoffs
enable the user to filter out noise by defining the level below which
milliamp outputs will indicate zero flow.
If a milliamp output indicates mass flow or volume flow, the user can
define a flow cutoff for the output. A flow cutoff is the lowest flow rate at
which the milliamp output indicates nonzero flow. If the flow signal drops
below the flow cutoff, the output goes to the current level that indicates
zero flow.
Flow cutoffs for milliamp outputs remain unaffected by the flow cutoff for
the transmitter frequency/pulse output and internal totalizers. To assign
a flow cutoff to the frequency/pulse output and internal totalizers, use the
Configure Transmitter Variables dialog box.
Vista de pagina 126
1 2 ... 122 123 124 125 126 127 128 129 130 131 132 ... 213 214

Comentarios a estos manuales

Sin comentarios