158
Using ProLink
®
Software with Micro Motion
®
Transmitters
Test Menu
continued
Test point diagnostics for Version 3 RFT9739
Performing the
frequency/pulse output
test
To perform a frequency/pulse output test:
1. Open the Test menu, then choose Frequency/Pulse Output Test. The
dialog box shown in Test Frequency Outputs dialog box then
appears.
2. Enter the desired output level, in Hz, into the Set Output To text box.
For an RFT9739, enter any frequency from 0.1 to 15,000 Hz
For an IFT9701 or IFT9703 with software version 1.2 or higher,
enter any frequency from 0.01 to 7200 Hz
For an IFT9701 or IFT9703 with a software version lower
than 1.2, enter any frequency from 0.01 to 1200 Hz
For an RFT9712 or RFT9729, enter any frequency from 1 to
10,000 Hz
3. Choose Set, then check the readout from the frequency counter. It
should indicate the same frequency as the frequency entered in the
Set Output To text box in step 2.
Figure 15-2. Test Frequency Outputs dialog box
15.4 Test point diagnostics for
Version 3 RFT9739
The ProLink program provides on-line test point information for Version 3
RFT9739 transmitters. Use the Test Points dialog box to view sensor
tube frequency, pickoff voltages, drive gain voltage, and live zero. Open
the Test menu, then choose Test Points. The Test Points dialog box
appears, as depicted in
Figure 15-3
, page 159.