Using ProLink
®
Software with Micro Motion
®
Transmitters
163
Applications Menu
continued
Application builder
Applications Menu Help Menu
Configure Menu:
Meter Factors
Calibrate Menu Test Menu
RFT9739
Choose Start to start the internal totalizers.
Choose Stop to stop the internal totalizers.
Choose Reset to reset the mass and volume total to 0.00.
To exit the Totalizer Control, double-click the control-menu box in the
upper left corner of the dialog box.
Totalizer functions can be disabled, depending on the RFT9739 security
mode. See
Table 16-1
. For more information about security modes, refer
to the RFT9739 instruction manual.
IFT9701/IFT9703
Choose Reset to reset the mass or volume total to 0.00.
RFT9712/RFT9729
Choose Start to start the internal totalizers.
Choose Stop to stop the internal totalizers.
Choose Reset to reset the mass or volume total to 0.00.
16.3 Application builder
The Application Builder is a future enhancement to the ProLink software
program.
Table 16-1. Totalizer Control for the RFT9739
Function Performed with Mode 1 Mode 2 Mode 3 Mode 4 Mode 5 Mode 6 Mode 7 Mode 8
Totalizer
reset*,
no flow
Scroll and Reset
buttons/knobs
Disabled Disabled Disabled Disabled
HART or Modbus
device
Disabled Disabled Disabled
Totalizer
reset*/start/
stop, with flow
Scroll and Reset
buttons/knobs
Disabled Disabled Disabled Disabled Disabled Disabled Disabled
HART or Modbus
device
Disabled Disabled Disabled Disabled Disabled Disabled
*
Resetting the totalizer has no affect on the mass or volume inventory. For more information, refer to “Outputs” in the RFT9739
manual.