Using ProLink
®
Software with Micro Motion
®
Transmitters
161
Applications Menu Help Menu
Configure Menu:
Meter Factors
Calibrate Menu Test Menu
16 Applications Menu
16.1 Overview
The Applications menu enables the user to control the transmitter
internal totalizers. The Application Builder command is designed as an
enhancement to the ProLink software program.
16.2 Totalizer control
For an RFT9739, RFT9712 or RFT9729, the Totalizer Control command
enables the user to reset, start, or stop the internal totalizers. For an
IFT9701 or IFT9703, the IFT9701/9703 Totalizer Control command
allows the user to reset the totalizer.
For RFT9712 and RFT9729 transmitters only, the totalizer will start
immediately after it is reset to zero.
To control totalizers, open the Applications menu, then choose Totalizer
Control or IFT9701 Totalizer Control. The Totalizer Control dialog box
appears as depicted in:
Figure 16-1
for an RFT9739, RFT9712, or RFT9729
Figure 16-2
for an IFT9701 or IFT9703
CAUTION
When the totalizer is stopped, the frequency/pulse
output is disabled.
If the frequency/pulse output is used for process
control, failure to set control devices for manual
operation could affect process control.
Before stopping the totalizer, set process control
devices for manual operation.
To enable the frequency/pulse output, restart the
totalizer.