70
Using ProLink
®
Software with Micro Motion
®
Transmitters
View Menu: Status
continued
Informational indicators
Event indicators
With flow, density, temperature, pressure, or assigned to an RFT9739,
RFT9712, or RFT9729 event, Event 1 or Event 2 switches OFF or ON
whenever the process variable crosses the setpoint.
With mass total or volume total assigned to an event, Event 1 or Event 2
switches ON and OFF according to the low or high configuration of the
alarm.
• With a low alarm, Event 1 or Event 2 switches ON when the user
resets the totalizer.
• With a high alarm, Event 1 or Event 2 switches OFF when the user
resets the totalizer.
For more information about configuring events, see
Chapter 12
.
"Error Cleared"
Error Cleared switches ON for 1 second after other indicators in the
Status window have switched OFF. Error Cleared switches OFF when
the user closes the Status window.
"Power Reset"
Power Reset switches ON after a power failure, brownout, or power
cycle has interrupted operation of the transmitter. The transmitter has a
nonvolatile memory, which remains intact despite power interruptions.
"Security Breach"
Security Breach applies only to Version 3.0, 3.5, and 3.6 RFT9739
transmitters. Security Breach switches ON to indicate the transmitter
security mode has been changed from security mode 8. Clear the
message by re-entering security mode 8 or by performing a master
reset.
For more information about RFT9739 security modes, or to perform a
master reset, refer to the transmitter instruction manual.
"Display Readback Error"
Display Readback Error applies only to Version 3.0, 3.5, and 3.6
RFT9739 transmitters with a local display. Display Readback Error
switches ON to indicate the value written to the transmitter display was
not properly received by the transmitter display.
If the error does not clear itself within 60 seconds, cycle power to the
transmitter (turn power OFF, then ON) to disable the display. Contact the
factory to replace a faulty display.