Using ProLink
®
Software with Micro Motion
®
Transmitters
67
View Menu: Status
continued
Informational indicators
Configure Menu:
Transmitter Outputs
Configure Menu:
Transmitter Information
Configure Menu: EventsView Menu: Status
Configure Menu:
Characterize
Configure Menu:
Transmitter Variables
"Transmitter Initializing"
Transmitter Initializing switches ON to indicate transmitter self-calibration
in progress at start-up or after power cycling.
"Calibration In Progress"
Calibration in Progress switches ON if the user performs a density
calibration or auto zero.
• For information about density calibration, see
Section 14.4
,
page 138.
• For information about auto zero, see
Section 14.2
, page 133 and
Section 14.3
, page 136.
Zero indicators
Zero Too Noisy switches ON if mechanical noise prevents the transmitter
from setting an accurate zero flow offset during transmitter zeroing. If
auto zeroing fails, Calibration Failure switches ON in the Operational
frame of the Status window.
Zero Too High or Zero Too Low switches ON if flow is not completely
shut off during sensor zeroing, thereby causing the transmitter to
calculate a zero flow offset that is too high or too low to allow accurate
flow measurement.
For more information about transmitter zeroing, see
Section 14.2
,
page 133 and
Section 14.3
, page 136.