Using ProLink
®
Software with Micro Motion
®
Transmitters
125
Configure Menu: Transmitter Information
continued
Sensor database for RFT9739 and RFT9712/RFT9729
Configure Menu:
Transmitter Outputs
Configure Menu:
Transmitter Information
Configure Menu: EventsView Menu: Status
Configure Menu:
Characterize
Configure Menu:
Transmitter Variables
The burst mode can be enabled or disabled by the user with the
appropriate button in the ProLink Transmitter Information dialog box. The
burst option (Burst On/Off) cannot be changed during off-line
configuration.
Table 11-2. HART commands for burst mode
11.5 Sensor database for
RFT9739 and
RFT9712/RFT9729
The Sensor Database describes the sensor that operates with the
RFT9739, RFT9712, or RFT9729. The user can change the sensor
database without affecting transmitter outputs or the flowmeter
configuration.
Serial #
Enter the serial number from the serial number tag attached to the
sensor housing.
Type
Enter a description of the sensor type (such as CMF200M for an ELITE
CMF200 sensor with 316L stainless steel flow tubes).
Flange, material, and liner
Open the list boxes in the Sensor Database frame to select from options
that describe the sensor flange type, flow tube construction material,
and flow tube liner material (see
Table 11-3
). Default options are shown
if the transmitter has not been configured. The last-chosen option is
shown if the transmitter has been configured.
Table 11-3. Flange types, tube and liner material options
HART command Description
1 Read primary variable
2 Read mA1 current and percent of range
3 Read dynamic variables and mA1 current
33 Read transmitter variables
110 Read all dynamic variables
Flange Material Liner
Sanitary ANSI 600 316L None
PN40 ANSI 900 Hastelloy-C PTFE
JIS 10K Union Monel Halar
ANSI150 SPECIAL Tantalum SPECIAL
ANSI 300 Unknown SPECIAL Unknown
Inconel*
Unknown
*The RFT9712 will not accept this setting