154
Using ProLink
®
Software with Micro Motion
®
Transmitters
Calibrate Menu
continued
Milliamp output trim
Figure 14-16.Milliamp output trim: enter measured high output
7. When the dialog box appears, choose Yes if the reference device
indicates approximately 20 mA. If the reference device indicates an
output level that is not acceptably close to 20 mA, choose No, then
repeat step 6.
8. After completing the milliamp output trim, choose Yes to enable the
output to produce a varying current.
If communication between the ProLink program and the transmitter is
interrupted before the trim procedure is complete, one of the following
indicators in the Status window switches ON:
• Analog 1 Fixed
• Analog 2 Fixed
• Analog Output Fixed
To view the Status window, open the View menu, then choose
Transmitter Status.
After checking indicators in the Status window, reopen the Calibrate
menu, then complete the milliamp output trim.