Density calibration
Table 14-3. Maximum flow rates for Micro Motion sensors
c. Use any established method to derive an accurate density, in
grams per cubic centimeter, for the fluid at line conditions. If water
is the high-density calibration fluid, a value from
Table 14-4
can
be used for the density.
d. Open the Calibrate menu, then choose Density. At the cascading
menu, choose Second Point (Water) to open the Density Point 2
dialog box, shown in
Figure 14-5
.
e. Enter the line-condition density, in grams per cubic centimeter,
into the Enter Density text box, then choose Do Cal. The
transmitter measures the tube period, corrects it to 0°C, and
stores the value. The indicator in the Density Point 2 dialog box
blinks ON and OFF while calibration is in progress. When the
indicator stops blinking, the high-density calibration is complete.
Sensor model
Maximum flow rate
lb/min kg/h
ELITE
®
CMF010 0.25 6.75
CMF025 5 135
CMF050 15 425
CMF100 62 1700
CMF200 200 5440
CMF300 625 17,010
CMF400 1250 34,090
BASIS
®
F025 5 135
F050 15 425
F100 62 1700
F200 200 5440
Model D
D6 0.125 3.25
D12 0.25 8.25
D25 1.5 42
D40 2.75 76
D65 18 510
D100 50 1360
D150 175 4760
D300 435 11,905
D600 1560 42,525
Model DH
DH6 0.125 3.25
DH12 0.25 8.25
DH25 1.5 42
DH38 3 85
DH100 50 1360
DH150 175 4760
DH300 435 11,905
Model DL
DL65 15 420
DL100 50 1360
DL200 215 5950
Model DT
DT65 18 510
DT100 50 1360
DT150 87 2380