74
Using ProLink
®
Software with Micro Motion
®
Transmitters
Configure Menu: Characterize
continued
Flow calibration factor
Table 8-1. Temperature coefficients for flow
Sensor model Coefficient Sensor model Coefficient
CMF010HP, high pressure 2.88
D sensor, stainless,
serial number equal to or below 214724 5.13
CMF010M, stainless 4.26
D sensor, stainless,
serial number above 214724 4.26
CMF010N, Inconel
®
2.79
D sensor, Hastelloy,
serial number equal to or below 214724 3.15
CMF025M, stainless 4.75
D sensor, Hastelloy,
serial number above 214724 2.79
CMF025H, Hastelloy
®
2.90
D600, serial number
equal to or below 214724 4.70
CMF050M, stainless 4.75 D600, serial number above 214724 4.50
CMF050H, Hastelloy 2.90 F025, stainless 4.65
CMF100M, stainless,
serial number below 357341 4.75 F050, stainless 4.77
CMF100M, stainless,
serial number equal to or above 357341 4.26 F100, stainless 4.32
CMF100H, Hastelloy,
serial number below 357341 2.90 F200, stainless 4.29
CMF100H, Hastelloy,
serial number equal to or above 357341 2.79 R025, stainless 4.65
CMF200M, stainless,
serial number below 332683 4.75 R050, stainless 4.50
CMF200M, stainless,
serial number equal to or above 332683 4.26 R100, stainless 4.67
CMF200H, Hastelloy,
serial number below 332683 2.90 DL025, Tantalum
®
2.18
CMF200H, stainless,
serial number equal to or above 332683 2.79 DS012, Tantalum 2.18
CMF300M, stainless,
serial number below 332692 4.75 DL050, Tantalum 2.18
CMF300M, stainless,
serial number equal to or above 332692 4.26 DX025, stainless 4.26
CMF300H, Hastelloy,
serial number below 332692 2.90
CMF300H, Hastelloy,
serial number equal to or above 332692 2.79
CMF300A 4.26
CMF400 3.89