Using ProLink
®
Software with Micro Motion
®
Transmitters
Configure Menu: Characterize
continued
Density factor for IFT9701/IFT9703 and RFT9712/RFT9729
Table 8-2. Methods for determining RFT9739 density factors
8.4 Density factor for
IFT9701/IFT9703 and
RFT9712/RFT9729
Density factors describe a particular sensor's density measurement
sensitivity. Testing conducted in the Micro Motion Flow Calibration
Laboratory determines the precise values of the density factors for each
sensor.
The IFT9701, IFT9703, RFT9712, and RFT9729 use a 13-digit density
calibration factor to calculate density (For an IFT9701, IFT9703,
RFT9712, or RFT9729 that is shipped without a sensor, the default value
— which
must
be changed — is 05000500004.44).
The first five digits
represent the natural tube period, in microseconds,
when the flow tubes contain a fluid (factory-calibrated with air) with a
density of 0.0000 g/cc. This value is also referred to as K1.
The second five digits
represent the natural tube period, in
microseconds, when the flow tubes contain a fluid (factory-calibrated
with water) with a density of 1.0000 g/cc. This value is also referred to as
K2.
Density
factor
Use the same method for all six density factors
Method 1 Method 2 Method 3
This method is preferred.
If possible, use this method.
This method is acceptable if
D1 and D2, or K1 and K2 are
not listed on the sensor tag.
This method is acceptable if Method 2
is not possible.
Dens A
Enter the
D1
value
from the sensor serial number
tag
Perform the density calibration
procedure, described in
Section 14.4
, page 138.
Enter the
D1
value from the sensor tag,
or enter 0.0012 if no
D1
value is listed
Dens B
Enter the
D2
value
from the sensor serial number
tag
Enter the
D2
value from the sensor tag,
or enter 0.998 if no
D2
value is listed
K1
Enter the
K1
value
from the sensor serial number
tag
Enter the first five digits of the 13-digit
density calibration factor from the sensor
serial number tag
K2
Enter the
K2
value
from the sensor serial number
tag
Enter the second five digits of the 13-digit
density calibration factor from the sensor
serial number tag
K3
Enter 0
(the factory-default value for K3)
Enter 0
(the factory-default value for K3)
FD
Enter the
FD
from the sensor tag
or calibration certificate
Contact Micro Motion for an appropriate
value for FD
Temp Coeff
Enter the
TC
value
from the sensor serial number
tag
Enter the Temp Coeff value
from Methods 1 or 3. If it is not
possible to read the sensor
tag, enter the appropriate
density temperature
coefficient from
Appendix C
.
Enter the last three digits and decimal
point from the 13-digit density calibration
factor from the sensor serial number tag