Using ProLink
®
Software with Micro Motion
®
Transmitters
View Menu: Status
continued
Fault outputs
7.2 Fault outputs
If a flowmeter failure occurs, transmitters produce fault outputs. Model
RFT9739, IFT9701, IFT9703, RFT9712, and RFT9729, produce
downscale or upscale outputs to indicate a fault. In addition, a Version 2
or earlier RFT9739 also has last measured value and internal zero fault
levels.
To set downscale or upscale fault output levels for a Version 3
RFT9739, an IFT9701, an IFT9703, an RFT9712, or an RFT9729,
refer to the transmitter instruction manual.
To set fault levels for a Version 2 or earlier RFT9739, open the
Configure menu, then choose Transmitter Outputs.
RFT9739
Downscale: milliamp outputs go to 0 mA if they produce a 0-20 mA
current, or to 0-2 mA if they produce a 4-20 mA current;
frequency/pulse output goes to 0 Hz.
Upscale: milliamp outputs go to 22-24 mA; frequency/pulse output
goes to 15-19 kHz.
Last measured value (not available with Version 3 transmitters):
Outputs hold the values measured immediately before the fault
condition occurred.
Internal zero (not available with Version 3 transmitters): Outputs
indicate a value of zero for the represented process variable.
IFT9701/IFT9703
Downscale: The milliamp output goes to 2 mA; the pulse output goes
to 0 Hz.
Upscale: The milliamp output goes to 22 mA; the pulse output goes
to 7200 Hz for version 1.2 and higher or 1200 Hz for lower versions.
RFT9712/RFT9729
Downscale: The milliamp output goes to 2 mA; the frequency/pulse
output goes to 0 Hz.
Upscale: The milliamp output goes to 22 mA; the frequency/pulse
output goes to 11,520 Hz.
7.3 Critical indicators
Critical indicators switch ON under the following conditions:
Master reset defaults in transmitter software
Transmitter failure
Flow, temperature, or density outside sensor limits
Faulty flowmeter wiring, cable, or connections
Inappropriate density calibration factors
Plugged or partially plugged flow tubes
Loss of input signal (when pressure is enabled)
Temperature calibration failed (RFT9712 and RFT9729 only)
Data loss during configuration