174
Using ProLink
®
Software with Micro Motion
®
Transmitters
Uploading and Downloading Configuration Files with a Model 268
continued
Downloading a configuration to a transmitter
B.2 Downloading a
configuration to a
transmitter
After the Model 268 has been uploaded, the configuration file can be
downloaded to a target transmitter.
Table B-1
, page 171 lists Model 268
software compatibility with Micro Motion transmitters.
To download the configuration file from the Model 268:
1. After uploading the Model 268, connect the Model 268 to the target
transmitter, then press RESTART.
2. Compare the measurement units in the target transmitter memory
with those in the source configuration file. The source configuration
must have the same measurement units as those in the target
transmitter memory. Otherwise, the target transmitter will refuse the
source configuration when it is downloaded.
a. After saving the work registers in the safe memory or proceeding
without saving the work registers, press REVIEW.
To review the measurement units from the source
configuration, press F2 (OFLN Mem).
To review the measurement units from the target transmitter,
press F3 (WORK Regs).
b. If necessary, enter the change input/output (Chng I/O) branch of
the Model 268 software, reconfigure measurement units so they
match the measurement units in the source configuration, then
send the units to the target transmitter.
3. After sending measurement units to the target transmitter, the user
can download the source configuration.
With the Model 268 still connected to the target transmitter, press
RESTART.
The Model 268 again uploads the work registers from the transmitter
memory.
4. After saving the work registers in the safe memory or proceeding
without saving the work registers, press F3 (Config) to access the
configuration branch.
CAUTION
Failure to enter calibration factors could cause
measurement error.
To avoid measurement error after downloading a
configuration file to a transmitter, use the Model 268 to
enter flow and density calibration factors into the
transmitter memory, or open the Configure menu and
characterize the sensor as described in Chapter 8.