Micro Motion RFT9709 Manual de instrucciones Pagina 77

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 214
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 76
Using ProLink
®
Software with Micro Motion
®
Transmitters
63
View Menu: Status
continued
Operational indicators
Configure Menu:
Transmitter Outputs
Configure Menu:
Transmitter Information
Configure Menu: EventsView Menu: Status
Configure Menu:
Characterize
Configure Menu:
Transmitter Variables
Table 7-7. Troubleshooting operational failures
"Slug Flow"
Programmed slug flow limits enable transmitter outputs and the ProLink
Status window to indicate conditions such as slug flow (gas slugs in a
liquid flow stream or liquid in a gas stream). Such conditions adversely
affect sensor performance by causing erratic vibration of the flow tubes,
which in turn causes the transmitter to produce inaccurate flow signals.
If the fluid density goes outside the user-defined slug flow limits, all of
the following occur:
The Slug Flow indicator in the ProLink Status window switches ON.
Milliamp outputs go to 0 mA if they produce a 0-20 mA current, or to
4 mA if they produce a 4-20 mA current.
Frequency/pulse output goes to 0 Hz.
Flow totalization stops while the density reading is outside density
limits.
The diagnostic LED on RFT9739 field-mount, IFT9701, IFT9703,
and RFT9712 transmitters blink OFF once per second (75% ON,
25% OFF).
The flowmeter resumes normal operation when liquid fills the flow
tubes (when density stabilizes within the programmed slug flow
limits).
Operational indicator Cause Corrective actions
Slug Flow Gas slugs causing process density to
go below low slug flow limit
Solids causing process density to go
above high slug flow limit
Open View menu, choose Process Variables
to monitor density
Open Configure menu, choose Characterize,
enter new slug flow limits
Open Configure menu, choose Transmitter
Outputs, enter new slug duration for Model
RFT9739
Frequency Saturated Flow rate driving RFT9739 output from
FREQ+ and RETURN terminals to 0 or
15 kHz
Flow rate driving RFT9712 or RFT9729
frequency/pulse output to 0 or 11,520
Hz
Open Configure menu, choose Transmitter
Variables, change flow measurement units
Open Configure menu, choose Transmitter
Outputs, rescale frequency/pulse output
Reduce flow rate
Frequency Output Saturated Flow rate driving IFT9701 or IFT9703
pulse output to 0 or 7200 Hz for versions
1.2 and higher, or 0 to 1200 Hz for lower
versions
Analog 1 Saturated RFT9739 output from PV+ and PV–
terminals is 0, 2, or 22 mA
RFT9712 or RFT9729 milliamp
output is 2 or 22 mA
Open Configure menu, choose Transmitter
Outputs, change value of variable at 20 mA
Alter fluid process
Analog 2 Saturated RFT9739 output from SV+ and SV
terminals is 0, 2, or 22 mA
Analog Output Saturated IFT9701 or IFT9703 milliamp output is 0
or 22 mA
Raw Flow Overflow
Raw Elec. Zero Overflow
Transmitter cannot calculate flow rate Eliminate pipe stress, vibration, or mechanical
noise
Vista de pagina 76
1 2 ... 72 73 74 75 76 77 78 79 80 81 82 ... 213 214

Comentarios a estos manuales

Sin comentarios